Model Number |
R.I 100 Series |
Refractive Index Detector Shodex RI-101, 102, 104
Differential refractive index detectors are widely used as general-purpose
detectors. However, the problem with
these detectors is the rather complicated start-up operations when compared with
other general-purpose detectors.
This is impact due to the limitation imposed by the detection principle by which
changes in the refractive index of
the eluent containing the sample are detected. Previous units were also difficult
to validate. The Shodex RI-101, our
latest detector, employs a number of improvements for shortening the time for
stabilization. By using a micro-processor
it is capable of automatic start-up from eluent replacement to checking the base
-line stabilization. This new concept
detector, Shodex RI-101, is a differential refractive index detector for use in
high performance liquid chromatography.
For the use with downsizing GPC columns, RI-104 is suited because of its small
flow cell volume. For semi-preparative
applications, RI-102 is the ideal detectors.
Features
- Exceptional ease - of - use:
(a) Equipped with color liquid crystal display, the detector allows the
chromatographer to easily see operational conditions and base line trends at a
glance.
(b) With the automatic start-up function, complicated operations from the
replacement of the eluent in the reference cell to checking of the base-line
stabilization can be made automatically.
(c) Easy validation and calibration of the unit assisted by the HELP function.
- Superior stability:
By optimizing the method of temperature control, the time for stabilization after
the power has been turned on has been shortened and superior base-line stability
has been achieved.
- Safety measures:
As the detector is equipped with a leak sensor, the pump can be set to stop
automatically in case of solvent leakage.
- Various automated functions:
The detector is fitted with external input and output terminals as well as RS232C
communication ports, permitting advanced automation by linking the detector with
various HPLC systems, enabling advanced automation by combining the detector with
various chromatographs.
Specifications
| Model |
RI-101
Analysis |
RI-104
Semi-micro |
RI-102
Preparative |
| Measuring method |
Deflection type |
| Refractive index range |
1/4 - 512uRIU |
1/4 - 512uRIU |
2.5 - 5120uRIU |
| Flow rate |
Ordinary |
0.2 to 3.0mL/min |
0.2 to 1.0mL/min |
1.0 to 50mL/min |
| Max. |
10mL/min (Eluent: Purified water) |
1mL/min (Eluent: Purified water) |
100mL/min (Eluent: Purified water)
|
| Linearity range |
600 RIU |
600uRIU |
6000uRIU |
| Noise level |
2.5nRIU (Response: 1.5sec) |
5nRIU (Response: 0.5sec) |
2.5nRIU (Response: 1.5sec) |
| Auto zero range |
Full Auto Zero |
| Auto zero resolution |
less than 1 (@ 8mV/u RIU) /4 (@
2mV/uRIU) n
RIU |
less than 1 (@ 8mV/u RIU) /4 (@ 2mV/u
RIU) n
RIU |
less than 1 (@ 0.8mV/u RIU) /4 (@
0.2mV/uRIU)
n RIU |
| Cell volume |
8micro-L |
2.5micro-L |
8micro-L |
| Dead volume |
IN to cell :ca. 60micro-L cell to
OUT:ca.600micro-L Total: ca.670micro-L |
IN to cell :ca. 10micro-L cell to
OUT:ca.355micro-L Total :ca.370micro-L |
IN to cell :ca.120micro-L cell to
OUT:ca.510micro-L Total: ca.640micro-L |
| Flow cell pressure |
49kPa(0.5kgf/cm2) |
| Integrator output |
0 to 1V (2mV/uRIU, 8mV/uRIU) |
0 to 1V (2mV/uRIU, 8mV/uRIU) |
0 to 1V (0.2mV/uRIU, 0.8mV/uRIU) |
| Recrder output |
0 to 10mV/FS |
| Contact output |
(1) Ready, (2) Solvent Leak (3)Error (Overheating,
Low Light Intensity, Null Glass Home Position Error, Lost
Parameters, Optical Balance ) (Contact Capacity: DC 24V 0.1
A.) |
| Contact input |
(1)Auto Zero, (2)Marker, (3)Purge
On/Off |
Operation support
functions |
(1)Automatic start-up (Start Up
Sequence) (2)Span / Validation Guide (3)Real Time Baseline
Monitor |
| External Communication |
RS232C |
| Response |
FAST,STD,SLOW |
| Temperature setting |
OFF, 30 to 50 deg-C(1deg-C increment), 77deg-C
Temp. Fuse |
| Solvent contact parts |
SUS316, Teflon, Quartz glass |
| Dimensions / Weight |
W260 x D400 x H200mm, ca.13kg |
| Power source |
AC100-240V -10% to 10%, 50/60Hz,
150VA |
| CE Mark Standards |
EN61326-1, EN61000-3-2, EN61000-3-3 and
EN61010-1 |
| Accessories |
Power cable, signal cable, tubing, fuse
etc. |
|
|
|
|
| |
|
|
|